Show simple item record

dc.contributor.advisorMilner, Thomas E.
dc.contributor.advisorRylander, H. Grady (Henry Grady), 1948-
dc.creatorYin, Biweien
dc.date.accessioned2014-10-15T20:55:31Zen
dc.date.issued2013-08en
dc.date.submittedAugust 2013en
dc.identifier.urihttp://hdl.handle.net/2152/26614en
dc.descriptiontexten
dc.description.abstractOptical Coherence Tomography (OCT) is a non-invasive three-dimensional imaging technique. OCT synthesizes a cross-sectional image from a series of lateral adjacent depth scans, and with a two-dimensional scanning scheme, three-dimensional intensity image of sample can be constructed. Due to its non-invasive capability, OCT has been widely applied in ophthalmology, cardiology and dermatology; and in addition to three-dimensional intensity image construction, various functional OCT imaging techniques have been developed for clinical application. My research is focused on developing functional OCT systems for application in ophthalmology, including polarization-sensitive optical coherence tomography (PS-OCT) for retinal nerve fiber layer (RNFL) birefringence measurement and dual-wavelength photothermal optical coherence tomography (DWP-OCT) for microvasculature blood oxygen saturation (SO2) measurement. In the study, a single-mode-fiber based polarization-sensitive swept-source OCT (PS-SS-OCT) with polarization modulator, polarization-sensitive bulk-optics balanced detection module is constructed and polarization processing methods based on Stokes vectors are applied to determine birefringence. PS-OCT is able to provide human subject's RNFL thickness, phase retardation, and birefringence information. Degradation in the degree of polarization (DOP) along depth is investigated and its difference between four quadrants of RNFL (superior, temporal, inferior and nasal) indicates the structural property difference. DWP-OCT is a novel functional OCT system consisting of a phase-sensitive optical coherence tomography system (PhS-OCT) and two photothermal excitation lasers. PhS-OCT is based on a swept-source laser operating in the 1060 nm wavelength range; the two photothermal excitation lasers with wavelength 770 nm and 800 nm are intensity modulated at different frequencies. PhS-OCT probe beam and two photothermal excitation beams are combined and incident on the sample, optical pathlength (op) change on the sample introduced by two photothermal excitation beams are measured and used for blood SO2 estimation. A polarization microscope is proposed for future study. The polarization microscope is an imaging technique providing molecular structure and orientation based on probe light's polarization state information. The polarization microscope uses a wavelength tunable light source, and can achieve any incident polarization state by a retarder-rotator combination. Specimen's birefringence can be determined based on the changing of detected light amplitude.en
dc.format.mimetypeapplication/pdfen
dc.subjectOptical coherence tomographyen
dc.subjectImagingen
dc.subjectBirefringenceen
dc.subjectPolarizationen
dc.subjectPhaseen
dc.subjectRetinal nerve fiber layeren
dc.subjectOxygen saturationen
dc.subjectMicroscopyen
dc.titleOptical coherence tomography for retinal diagnosticsen
dc.typeThesisen
dc.date.updated2014-10-15T20:55:31Zen
dc.description.departmentElectrical and Computer Engineeringen
thesis.degree.departmentElectrical and Computer Engineeringen
thesis.degree.disciplineElectrical and Computer Engineeringen
thesis.degree.grantorThe University of Texas at Austinen
thesis.degree.levelDoctoralen
thesis.degree.nameDoctor of Philosophyen